Reagent Purity and Reference Materials
Because modern laboratories are highly automated, most reagents are provided in ready-to-use "kits" rather than prepared from raw chemical stocks. However, understanding the grade of these chemicals is vital for quality control.
Grades of Purity
- Analytical Reagent (AR): The standard for most analytic laboratory procedures. The Committee of American Chemical Society (ACS) establishes the specifications for these chemicals.
- Ultrapure/Chemically Pure (CP): These undergo additional purification (e.g., chromatography) and are often designated as HPLC grade.
- USP and NF: These grades are used for drug manufacturing. Their purity is based on being non-injurious to humans rather than meeting specific analytical assay requirements.
- Technical/Commercial Grade: These should never be used in a clinical laboratory.
Standard Reference Materials (SRMs)
In clinical chemistry, many biological constituents cannot meet the extreme purity requirements of a "primary standard". To solve this, the National Institute of Standards and Technology (NIST) certifies Standard Reference Materials (SRMs). Manufacturers use these SRMs to produce calibrators and standards, ensuring that all laboratory results are traceable to a national standard.
Water Quality and Purification
Water is the most frequently used reagent in the laboratory, making its purity critical. Tap water is unsuitable for most procedures; instead, laboratories use purified reagent-grade water.
Common purification methods include:
- Distillation: Boiling water and condensing the vapor to remove organic materials and particulates.
- Deionization (Ion Exchange): Using resins to remove ions, replacing them with hydroxyl or hydrogen ions.
- Reverse Osmosis: Using pressure to force water through a semipermeable membrane.
- Ultrafiltration: Excellent for removing microorganisms and particulate matter.
The Clinical and Laboratory Standards Institute (CLSI) classifies reagent water into six categories based on application. A key indicator of water purity is resistance; as purity increases, electrical resistance increases linearly.
Concentration and Chemical Properties
Expressing Concentration
Concentration describes the amount of solute in a solution. While not all of these are SI units, they are fundamental to clinical practice:
- Percent Solution:
- Weight/Weight (w/w): grams of solute/100g of solution
- Volume/Volume (v/v): mL of solute/100mL of solution
- Weight/Volume (w/v): grams of solute/100mL of solution (Common in clinical labs).
- Molarity (M): moles of solute/1L of solution.
- Molality (m): moles of solute/1kg of solvent. Unlike molarity, molality is independent of temperature and pressure changes.
- Normality (N): gram equivalent weights/1L of solution.
Colligative Properties and Osmolality
Colligative properties depend on the number of solute particles, not their identity. These include osmotic pressure, vapor pressure, and freezing point depression.
In clinical settings, osmolality is preferred over osmolarity. Because osmolality is based on the mass of the solvent (molality), it remains stable regardless of changes in temperature or pressure.
Redox, pH, and Buffers
- Redox Potential: Measures a solution's ability to accept (oxidizing agent) or donate (reducing agent) electrons. A helpful mnemonic is LEO (Lose Electrons, Oxidized) and GER (Gain Electrons, Reduced).
- pH and Buffers: pH is the negative logarithm of hydrogen ion concentration:
pH=−log[H+].
Buffers are weak acids or bases that minimize changes in pH. The Henderson-Hasselbalch equation describes the relationship between pH, $pK_a$, and the ratio of the conjugate base to the weak acid:
pH=pKa+log[HA][A−].
Laboratory Equipment and Calibration
Temperature Measurement
All analytic reactions occur at an optimal temperature. For enzyme determinations, precise control is mandatory.
- Liquid-in-Glass Thermometers: Use colored liquid; must be checked for bubbles or separation.
- Electronic Thermistors: Small, with millisecond response times; commonly used in automated instruments.
- Calibration: All devices must be calibrated against NIST-certified or traceable thermometers. The SI unit for temperature is Kelvin (K).
Glassware and Plasticware
Laboratory vessels are classified by their precision (ASTM standards):
- Class A: High precision; preferred for clinical applications.
- Class B: Lower precision; often used in student laboratories.
Vessels are also categorized by how they hold liquid:
- To Contain (TC): Holds a specific volume but will not dispense that exact amount when poured.
- To Deliver (TD): Designed to dispense the exact volume indicated.
Pipettes and Balances
Pipettes are used to transfer volumes, typically under 20 mL.
- Volumetric Pipettes: Highly accurate, self-draining, and preferred for dilutions.
- Automatic Pipettes: Include air-displacement (uses disposable tips) and positive-displacement (piston moves within the tip).
For weighing, analytical balances are used for high-precision tasks like preparing primary standards, while electronic balances offer fast response times via electromagnetic force.
Centrifugation
Centrifuges separate components based on mass and density. While speed is measured in revolutions per minute (rpm), the clinical standard is Relative Centrifugal Force (RCF), also known as "g".
RCF=1.118×10−5×r×(rpm)2
Exam Tip: Centrifugation Safety
Always balance a centrifuge by placing tubes of equal volume and weight in opposition. An unbalanced rotor causes vibration and noise, which can damage the equipment and lead to errors in separation.
Laboratory Mathematics: Dilutions and Beer's Law
Dilutions
A dilution expresses the concentration of a substance relative to the total final volume. When a result exceeds an assay's linearity, the sample must be diluted. To find the final result, multiply the measured value by the dilution factor (the reciprocal of the dilution).
Serial Dilutions involve a series of progressive dilutions where each step is prepared from the preceding one. This is useful when sample volume is limited or when determining a titer.
Beer-Lambert Law
The Beer-Lambert law is the foundation of spectrophotometry, stating that absorbance (A) is directly proportional to analyte concentration (c):
$A = abc$
Where a is the absorptivity constant and b is the light path length.
For enzyme assays, the change in absorbance over time (ΔA) is used to calculate concentration based on molar absorptivity (ϵ):
C=SVϵΔA⋅10−6TV
(Note: SV = sample volume, TV = total volume).